PN# R116P40
Description
Maxtec Max-18 Oxygen Sensor
Cross Reference
- Analytical Industries: PSR-11-33-1, PSR-11-33-2
- Ceramatec: CAG-18
- Envitec: OOM107-2
- Hudson RCI: V-02, 5569, 5525, 5528, 5529, 5550, 5552
- Integral Process: 65000
- MSA: 485905, 655262
- Maxtec: Max-18 (R116P40)
- Megamed: M1
- Mercury Medical: 10-103-01
- Nuova: E-33/1
- Oxitron: 3050
- Sensidyne: SV-02
- Sensoronics; Pacifitech: PT-02
- SensorTech: ST-02
- Teledyne: C1R, #B-39022, C2R
- Teleflex (Hudson): 5569, V-02, 5525, 5528, 5529, 5550, 5552
- Ventronics: 5525, 5528, 5529
- Viamed: C-2R
Specifications
- Range of Measurement: 0 to 100% O2
- Output Range: 200 ± 40 µA
- Zero Offset: < 5 µA
- Response Time: ≤ 13 s
- Linearity: < 3%
- Expected Lifetime: > 250,000 %O2 hours
- Operational Temperature Range: 0° to 40°C
- Storage Temperature Range: -20° to 50°C
- Humidity Range: Up to 100% RH
- Ambient Pressure Range: 600-1750 mbar
- Interference: Per EN ISO 80601-2-55
- Electrical Interface: 1mm Dia. x 3.5mm Long Pins. White Negative, Black Positive