PN# R115P10
Description
Maxtec Max-13 Oxygen Sensor
Cross Reference
- Analytical Industries: PSR-11-917-J
- City Technology: MOX-2
- Covidien: 4-020933-00, 7820 Monitor, 7200 Vent
- Datascope: Anestar S
- EnviteC: OOM-102-1
- HillRom: 6735142
- Invivo: 9445 Sensor
- Lifecare: PLV102
- MEDIX: Natalcare
- MSA: 406931, MiniOX 3000, MiniOx I, MiniOX II, MiniOX III
- Mercury Medical: 10-103-05
- Pacifitech: PT-11A
- Philips (Respironics): Criterion Vent, Espirit Vent Lifecare (#PLV-102), 37-079 Sensor, V200
- Precision Medical: PM5900, PM5900L, 504877
- Puritan Bennet: 7820
- Stephan: Christina
- Taema: Ceasar
- Teledyne: R17MED, #C-43690, R17MED, #C-43690, AX 300, MX300
- Teleflex (Hudson): 5567, V-11
- VTI: 100580
- Viasys: 66030 Sensor, 6400 Vent, 8400 Vent, VIP Vent
Specifications
- Range of Measurement: 0 to 100%
- Output Range: 13.0 to 16.5 mV
- Zero Offset: ≤ 0.25 mV
- Response Time: ≤ 12 s
- Linearity: ≤ 3%
- Expected Lifetime: > 1,000,000 %O2 hours
- Temperature Compensation: NTC
- Operational Temperature Range: 10° to 40°C
- Storage Temperature Range: -20° to 50°C
- Humidity Range: 0% to 99% RH, non-condensing
- Ambient Pressure Range: 700-1250 mbar
- Interference: PER EN ISO 80601-2-55
- Electrical Interface: 3.5mm mono jack