PN# R113P10
Description
Maxtec Max-11 Oxygen Sensor
Cross Reference
- Analytical Industries: PSR-11-915-2, PSR-11-915-2i, G0-080 (Old CnS SKU)
- Draeger: 6803290, 6850645, Narkomed G5, Cato, Narkomed 6000, Narkomed 6400, Anemone, Babylog 8000, Evita 2, Evita 2 Dura, Evita 4, Evita Series, Fabius, Fabius GS, Incubator 8000, Narkomed 1A, Narkomed 2, Narkomed 2B, Narkomed 3, Narkomed 4, Narkomed GS, Oxycom, Oxycom 100 D, Oxydig, PM8030, SA2
- EnviteC: OOM-201, OOM-104
- GE (Datex Ohmeda): 6051-000-224
- Hamilton Medical: HM-14, HM-13
- HillRom: PSR-11-915-2
- International Technologies: M-02
- Invivo: Magnitude
- Maxtec: Max-11 (R113P10), Max-11i (R113P06)
- Mercury Medical: 10-103-07
- MSA: 655263, 655264
- Nuova: E-15/2, E-8000
- Pacifitech: PT-14A
- Teledyne: R-23 MED, #C-61139
- Teleflex (Hudson): 5557, V-14
- Viamed: R-23V
- VTI: 101372, 100352, 101372, 100352
Specifications
- Range of Measurement: 0 to 100%
- Output Range: 14 to 20 mV
- Zero Offset: ≤ 0.20 mV
- Response Time: ≤ 12 s
- Linearity: ≤ 3%
- Expected Lifetime: > 500,000 %O2 hours
- Temperature Compensation: none
- Operational Temperature Range: 10˚ to 40˚C
- Storage Temperature Range: -20˚ to 50˚C
- Humidity Range: 0% to 99% RH, non-condensing
- Ambient Pressure Range: 700-1250 mbar
- Interference: PER EN ISO 80601-2-55
- Electrical Interface: gold-plated slip ring