PN# R113P90
Description
Maxtec Max-1 replacement oxygen sensor for GE (Datex Ohmeda) 0237-2034-700 Excel Modulus,5250 RGM,5150 Handheld Monitor, GE (Ohmeda) 5120, 5100, 5250 RGM, 4700 Oxicap, 7800, 5125, 5150, Handheld Monitor, 0237-2034-700, Excel Modulus, Analytical Industries PSR-11-915, Datascope Multinex Analyzer, Hewlet Packard 15201A, Mercury Medical 10-103-07, Pacifitech PT-10, Teledyne R15 #C-41131, Teleflex (Hudson) 5556, VTI 100352, 101272, 100352
Cross Reference
- Analytical Industries: PSR-11-915
- Biotest: S101
- CR: R-15
- Ceramatec: CAG-1
- City Technology: MOX-6
- Datascope: Multinex Analyzer
- Draeger; Air Shields: C-100
- Draeger: 6803290
- Envitec: OOM101
- GE (Datex-Ohmeda): 0237-2034-700, Excel Modulus, 5250 RGM, 5150 Handheld Monitor, 125, 205, 51, 510, 5100, 5120, 5125, 5140, 5200, 5250, 5400, 7200, 7800, 7810, Excel 210-7800, Excel 7800, Modulus 4700, Modulus 7800, Modulus II 7810, Oxicap, S5172802, 5250 RGM
- GE (Ohmeda): 5120, 5100, 5250 RGM, 4700 Oxicap, 7800, 5125, 5150, Handheld Monitor, 0237-2034-700, Excel Modulus
- Hamilton Medical: HM-10, HM-13
- Hewlett Packard: 15201A
- Hudson RCI: 5556
- Integral Process: 65001
- Maxtec: Max-1 (R113P90)
- Mercury Medical: 10-103-07
- MSA: 655264, 655263
- Nuova: E-15/0
- Pacifitech: PT-10
- Philips: 15201A
- Sensidyne: SV-14A
- SensorTech: ST-10
- Sensoronics; Pacifitech: SS-10A, PT-10
- Teledyne: R15 #C-41131
- Teleflex (Hudson): 5556
- VTI: 100352, 101272, VT1272
Specifications
- Range of Measurement: 0 to 100%
- Output Range: 14 to 20 mV
- Zero Offset: ≤ 0.30 mV
- Response Time: ≤ 17 seconds
- Linearity: ≤ 3% full scale
- Expected Lifetime: > 500,000 %O2 hours
- Temperature Compensation: none
- Operational Temperature Range: 10˚ to 40˚C
- Storage Temperature Range: -20˚ to 50˚C
- Humidity Range: 0% to 90% RH, non-condensing
- Ambient Pressure Range: 700-1250 mbar
- Interference: per EN ISO 80601-2-55
- Electrical Interface: gold-plated slip rings